A Jouyban, J Ye, B J Clark
Index: Pharmazie 57(4) , 248-9, (2002)
Full Text: HTML
A non-aqueous capillary electrophoresis method has been proposed to analyse hexetidine and its degradation products in a commercial liquid pharmaceutical formulation. The separation was achieved using a 37 cm fused silica capillary with a 75 microns internal diameter, an electrolyte of 120 mM sodium acetate buffer in methanol, at 16 kV and 25 degrees C. The results show that the proposed method is able to resolve hexetidine from its degradation products and/or impurities.
| Structure | Name/CAS No. | Molecular Formula | Articles |
|---|---|---|---|
![]() |
Hexetidine
CAS:141-94-6 |
C21H45N3 |
|
Physicochemical characterization of hexetidine-impregnated e...
2002-06-01 [Pharm. Res. 19(6) , 818-24, (2002)] |
|
The effect of hexetidine mouthwash on the prevention of plaq...
2011-08-01 [Int. J. Dent. Hyg. 9(3) , 182-90, (2011)] |
|
The plaque inhibitory effect of a CPC mouthrinse in a 3-day ...
2010-11-01 [Int. J. Dent. Hyg. 8(4) , 269-75, (2010)] |
|
Fluconazole and/or hexetidine for management of oral candidi...
2005-09-01 [Oral Dis. 11(5) , 309-13, (2005)] |
|
Parotid gland swelling following mouthrinse use.
2010-11-01 [Int. J. Dent. Hyg. 8(4) , 276-9, (2010)] |