抗抑郁药2/精神兴奋药/长期血浆质控品LC-MS/MS

询价
Chromsystems
德国
2021-07-15 21:50

上海谱芬生物科技有限公司

我要认领
上海谱芬生物科技有限公司
肖经理
13636535930
zhihong.xiao@cmass.com.cn
946433651
产品属性
CAS号药物质控品
供应商上海谱芬生物科技有限公司
数量8
注册证号详见COA
国食药监械注册号详见COA
英文名MassCheck® Antidepressants 2/ Psychostimulants/EXTENDED Plasma Controls
保质期详见COA
保存条件-18°
规格5 x 1 ml;2 x5 x 1 ml;2 x 5 x 1 ml
产品说明

联系方式:13636535930.www.cmass.com.cn
MassCheck® Antidepressants 2/ Psychostimulants/EXTENDED Plasma Controls

  • Atomoxetine
  • Bupropion
  • Clomethiazole
  • Dosulepin
  • erythro-Dihydrobupropion
  • Hydroxybupropion
  • Methylphenidate
  • Mianserin
  • Milnacipran
  • Moclobemide
  • N-Desmethyldosulepin
  • Opipramol
  • Reboxetine
  • Ritalinic acid
  • threo-Dihydrobupropion
  • Tianeptine
  • Tranylcypromine
  • Trazodone
  • Vilazodone
  • Vortioxetine

MassTox® TDM Series A

DESCRIPTION

MassCheck® Antidepressants 2/Psychostimulants/EXTENDED Plasma Controls, lyophilised.
For MassTox® TDM Series A Parameter Set Antidepressants 2/Psychostimulants/EXTENDED in Serum/Plasma.

CMS-CH0230/XT MassCheck® Antidepressants 2/Psychostimulants/EXTENDED Plasma Control, Bi-Level I + II (lyoph.), 2 x 5 x 1 ml
CMS-CH0231/XT MassCheck® Antidepressants 2/Psychostimulants/EXTENDED Plasma Control, Level I (lyoph.), 5 x 1 ml
CMS-CH0232/XT MassCheck® Antidepressants 2/Psychostimulants/EXTENDED Plasma Control, Level II (lyoph.), 5 x 1 ml

These lyophilised quality controls from Chromsystems are based on human plasma and should be used according to the instruction manual. They are designed to monitor the accuracy and precision of analytical procedures in clinical diagnostics for the quantitative determination of antidepressants 2 and psychostimulants in serum/plasma. While following the instructions the controls should be handled and measured in the same manner as a patient specimen.

TECHNICAL DATA

Method of Analysis

LC-MS/MS

Parameter

Atomoxetine, Bupropion, Clomethiazole, Dosulepin, erythro-Dihydrobupropion, Hydroxybupropion, Methylphenidate, Mianserin, Milnacipran, Moclobemide, N-Desmethyldosulepin, Opipramol, Reboxetine, Ritalinic acid, threo-Dihydrobupropion, Tianeptine, Tranylcypromine, Trazodone, Vilazodone, Vortioxetine

Concentrations

Please note: Concentration values shown here are typical. For lot specific concentration values please refer to the product insert supplied with the materials. The product inserts can also be requested from us directly or downloaded from our website after login.

SubstanceUnitTarget value
Level I
Target range
Level I
Target value
Level II
Target range
Level II
Atomoxetineµg/l129103 -155890712 -1068
Clomethiazolµg/l620496 -74446733738 -5608
N-Desmethyldosulepinµg/l83.967.1 -101215172 -258
Dosulepinµg/l49.039.2 -58.8339271 -407
Methylphenidateµg/l10.38.24 -12.457.445.9 -68.9
Mianserinµg/l23.919.1 -28.7128102 -154
Milnacipranµg/l58.346.6 -70336269 -403
Moclobemideµg/l389311 -46721851748 -2622
Opipramolµg/l91.773.4 -110541433 -649
Reboxetineµg/l56.044.8 -67.2297238 -356
Ritalinic acidµg/l61.449.1 -73.7255204 -306
Tianeptineµg/l54.243.4 -65326261 -391
Tranylcypromineµg/l69.055.2 -82.8561449 -673
Trazodoneµg/l515412 -61821651732 -2598
Vilazodoneµg/l29.823.8 -35.886.168.9 -103
Vortioxetineµg/l15.812.6 -1983.466.7 -100
Bupropionµg/l35.928.7 -43.1272218 -326
erythro-Dihydrobupropionµg/l34.527.6 -41.4261209 -313
Hydroxybupropionµg/l234187 -28118751500 -2250
threo-Dihydrobupropionµg/l178142 -21412511001 -1501

Reconstitution

Carefully remove the metal seal and rubber stopper from the vials. Add exactly 1.0 ml of distilled water into each vial. Return the stopper and incubate for 10 to 15 minutes at room temperature. Swirl the vials to dissolve the contents until homogeneity.

Storage conditions

Stored below -18 °C and unopened, the lyophilised controls are stable until the expiry date specified on the vial labels. The reconstituted controls can be stored up to 1 day light protected and tightly capped at +2 to +8 C. If the controls cannot be used within this period, they should be aliquoted and stored frozen below -18 °C (up to 3 months). For more details about storage conditions please check your instruction manual.