A test method to evaluate chronic particulate generation from durable polymer stent coatings: a case study of CYPHER® sirolimus-eluting coronary stents†
Judith H. Cohen,Karin M. Balss,Eugena Akerman-Revis,Mark Inderbitzen,James Lisk,Asha Marathe,Robert O'Brien,Elaine Pottinger-Cooper,William Rudd,Cheryl Wendel,George Papandreou,Cynthia A. Maryanoff
Analytical Methods Pub Date : 09/21/2012 00:00:00 , DOI:10.1039/C2AY25030F
Abstract

In order to predict the long term coating durability of drug-eluting stent (DES) products, we developed a new testing methodology using a novel mechanical tester to evaluate the potential of drugpolymer coatings to generate particles over time. The tester simulates the device radial fatigue expected during in vivo conditions due to expansion and contraction of the implanted vessel. Analysis by Scanning Electron Microscopy (SEM) of particulate collected via an in-line filtration system allows for evaluation of the chronic durability of DES polymer coatings. Data provided within show the durable polymer coating of CYPHER® Stents generates small amounts of particulates upon fatigue indicating coating integrity is maintained even when the DES system is subjected to expansion, deployment, and repetitive cycling modalities.

Graphical abstract: A test method to evaluate chronic particulate generation from durable polymer stent coatings: a case study of CYPHER® sirolimus-eluting coronary stents